Single Inductively Coupled Plasma-Atomic Emission Spectrometry-DW

Laboratory Fee Schedule
Group: Inorganic
Procedure #:  EZZ0195A
 

Synonym(s):  Metals in Water
Requisition Form G-22
Test Description This standard operating procedure describes the daily operation and analysis procedures for the determination of selected elements in drinking water samples by EPA Method 200.7.
Pre-Approval Needed N/A
Supplemental Information Required N/A
Supplemental Form(s) N/A
Performed on Specimens from (sources) Environmental
Sample/Specimen Type for Testing Drinking water 
Minimum Volume/Size Required 1 Liter
Storage/Preservation Prior to Shipping Store in hard or soft plastic, or glass; 1 quart (1 L)
Transport Medium N/A
Specimen Labeling
  • Label each bottle with a waterproof marker.
  • Sample identification must match information on submission form.
Shipping and Specimen Handling Requirements Ship according to Dangerous Goods Regulations, IATA, and/or CFR 49.
Turn-around Time
  • Sample must be analyzed within 180 days of sample collection. 
  • The target sample turnaround time is 14 days. 
Interferences/Limitations Some known interferences are as follows:
  • Spectral interference
  • Background emission and stray light
  • Spectral overlaps
  • Physical, chemical and memory interferences
Common Causes for Rejection
  • Insufficient quantity
  • unacceptable specimen type or source
  • improper shipping conditions
  • expired media or collection container
  • discrepancies between specimen label and submission form
  • incomplete or missing submission form
Additional Information Contact laboratory for supported elements.